3.2 Remote control commands
751 GPD Titrino
157
3.2.2.
3.2.2.70.
70. M
Mode.P
.Parameter.C
.Calibration.S
.SignalDrift
0.5...22...999, OFF
M
Mode.P
.Parameter.C
.Calibration.E
.EquTime
0...110
110...9999, OFF
Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding crite-
rion is switched off. If both criterions are on OFF, the measured value is ac-
quired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 20. After it has been edited once, it
remains in force with the set value.
3.2.2.
3.2.2.71.
71. M
Mode.P
.Parameter.C
.Calibration.El
.ElectrodeId
up to 8 ASCII char
Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.131.
3.2.2.
3.2.2.72.
72. M
Mode.P
.Parameter.C
.Calibration.Sm
.SmplChanger
ON, OFF
OFF
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.
3.2.2.73.
73. M
Mode.P
.Parameter.C
.Calibration.A
.ActPulse
first, all, OFF
OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page 224.
3.2.2.
3.2.2.74.
74. M
Mode.P
.Parameter.S
.Sequence.X.S
.X.Select
method, pause, L4 output,
L6 output, info, prep, stirrer, OFF
OFF
Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.75.
3.2.2.
3.2.2.75.
75. M
Mode.P
.Parameter.S
.Sequence.X.M
.X.Method
Method name
M
Mode.P
.Parameter.S
.Sequence.X.P
.X.Pause
00...999 999, INF
M
Mode.P
.Parameter.S
.Sequence.X.L
.X.L4Output
active, inactive,
pulse, OFF
OFF
M
Mode.P
.Parameter.S
.Sequence.X.L6
.X.L6Output
as for L4
M
Mode.P
.Parameter.S
.Sequence.X.I
.X.Info
up to 16 ASCII characters
M
Mode.P
.Parameter.S
.Sequence.X.Pr
.X.Prep
internal D0
internal D0, external D1,
external D2
M
Mode.P
.Parameter.S
.Sequence.X.S
.X.Stirrer
ON
ON, OFF
Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the
card. Up to 8 ASCII characters.
.Pause:
Pause time in s. INF means infinite. Continue the sequence with &m
$G.
.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin