3.2. Remote control commands
751 GPD Titrino
146
M
Mode.P
.Parameter.T
.TitrPara.St
.StartV.V
.V
00...999.99
M
Mode.P
.Parameter.T
.TitrPara.St
.StartV.F
.Factor
00...±999 999
M
Mode.P
.Parameter.T
.TitrPara.St
.StartV.R
.Rate
0.01...150, max.
max.
Parameters for DET, MET, SET, KFT, STAT: Start volume.
If an absolute start volume (abs.) has been selected, the volume in mL is valid.
A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size
*
factor
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.
3.2.2.16.
16. M
Mode.P
.Parameter.T
.TitrPara.P
.Pause
00...999 999
Parameters for DET, MET, SET, KFT, STAT: Pause time in s. Is waited off after
the dispensing of the start volume.
3.2.2.
3.2.2.17.
17. M
Mode.P
.Parameter.T
.TitrPara.DosU
.DosUnit
internal D0
internal D0, external D1/D2
M
Mode.P
.Parameter.D
.DosPara.DosU
.DosUnit
internal D0
internal D0, external D1/D2
.TitrPara.DosUnit: Parameter for DET, MET, SET, KFT, STAT. Selection of dos-
ing unit.
.DosPara.DosUnit: Parameter for DOS, DOC. Selection of dosing unit.
3.2.2.
3.2.2.18.
18. M
Mode.P
.Parameter.T
.TitrPara.Me
.MeasInput
11, 2, diff.
M
Mode.P
.Parameter.T
.TitrPara.I.Ipol
-127...11...+127
M
Mode.P
.Parameter.T
.TitrPara.Up
.Upol
-1270...400
400...+1270
M
Mode.P
.Parameter.T
.TitrPara.Po
.PolElectrTest
ON, OFF
OFF
Parameters for DET, MET, SET, KFT STAT:
Selection of the measuring input; valid with measured quantities pH and U.
"diff." means differential amplifier, see page 215.
With Ipol, the inquiries for the polarization current in uA (Ipol) and .PolElectrTest
are valid.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in
steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on changeover
from the inactive state to an active state (titration or conditioning).
3.2.2.
3.2.2.19.
19. M
Mode.P
.Parameter.T
.TitrPara.T
.Temp
-170.0...25.0
25.0...500.0
Parameters for DET, MET, SET, KFT, STAT: Titration temperature in °C. If a
Pt100 or Pt1000 is connected, the temperature is measured continuously and
the parameter .Temp is updated.
The temperature is used for the temperature correction in pH measurements.
3.2.2.
3.2.2.20.
20. M
Mode.P
.Parameter.St
.StopCond.V
.VStop.T
.Type
abs.
abs., rel., OFF
M
Mode.P
.Parameter.St
.StopCond.V
.VStop.V
.V
0...99.99
99.99...9999.99
M
Mode.P
.Parameter.St
.StopCond.V
.VStop.F
.Factor
0...
±
999 999
999 999