Metrohm 751 GPD Titrino User Manual
Page 59

2.6. Parameters, key
751 GPD Titrino
56
temperature 25.0 °C
Temperature (-170.0...500.0 °C)
If a T sensor is connected, the temperature is
measured continuously and the pH values are
corrected accordingly.
The last measured value is entered as the
"temperature" parameter.
Without a T sensor, the temperature entered manually
applies.
>stop conditions
stop V: OFF
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
filling rate max. ml/min
**titr.
Stop conditions
If a stop is not made after the dispensing time or after
the dispensed volume.
Type of stop volume (abs., rel., OFF)
"abs":
absolute stop volume in mL
"rel.":
relative stop volume to sample
"OFF":
stop volume switched off, not
monitored.
If "abs." is set:
Absolute stop volume (0...99 999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...± 999 999)
Calculated as: Stop V in mL = factor * sample size
Filling rate (0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
>monitoring
meas.mode: OFF
**titr.
meas.input: 1
Monitoring of measured values
Limit value violations are marked in the measuring
point list.
Measured values and temperatures are only stored in
the measuring point list if their monitoring is active.
Measured values (pH, U, Ipol, Upol, OFF)
If
≠
"OFF" follow the requests:
Measuring input (1, 2, diff.)
Request only with measuring modes pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 215.