4.1. Error and special messages
751 GPD Titrino
190
rate too low
With DOS the calculated rate is too low. It can not be achieved
with the mounted Exchange Unit.
Exit:
volume or change dosing conditions.
rates not corresponding
With STAT the number of rates that could be calculated is not
equal to the number of time windows set or the mean rate C80
could not be calculated. The calculation of a rate requires at
least 4 points.
result out of limits
The result lies outside the limits which were defined in the
method, see page 75.
Exit: Calculate result again or new start.
same buffer
In the calibration sequence the voltage difference between the
first and second buffer is < 6 mV.
Exit:
calibration).
sample size out
The sample size is outside the limits which are defined in the
method, see page 19.
Exit: Enter new sample size.
save lines OFF
The function "save lines" is not active although a submethod of
TIP contains assignments to C24 or C25.
Exit: DATA> key. Warning: The results of this line will not be saved. second TIP call In TIP no further TIP can be called up as a submethod. Exit: service is due The service interval has elapsed. Contact Metrohm service so silo empty The silo memory is switched in but empty and a titration has been started. Corrective action: At least the first 1 silo line before starting the first titration. Exit: silo full The silo memory is full up. Corrective action: If you have filled less than 255 silo lines, you can create more space by deleting old methods no longer needed. 1 silo line needs 18...120 bytes. Exit: stop EP reached A DET or MET titration was stopped as the stop criterion "stop EP" was reached. stop meas.val.reached A DET or MET titration was stopped as the stop measuring value pH, U or I was reached. stop rate reached STAT has been stopped as the stop rate has been reached. stop time reached SET or STAT has been stopped as the stop time has been reached. stop V reached The determination has been stopped as the stop volume has been reached. system error 3 The instrument adjustment data have been overwritten. Exit: message appears each time the instrument is switched on until it has been readjusted (Metrohm service).
that the Titrino can be serviced. This message will appear
each time the Titrino is switched on.
Exit: New start.