2.12. TIP, Titration procedure
751 GPD Titrino
86
Information for the commands (steps):
Command
Meaning
Input range
method
Method from the user memory or from the card. This
method runs as a submethod.
Name
pause
Waiting time. The waiting time can be aborted with
0...999 999 s,
inf.
L4, L6 output
Set L4 output (pin 3) resp. L6 output (pin 1) of the
remote socket.
active = 0 V, inactive = 5 V, pulse > 100 ms, off =
output is not used.
Cable Titrino (L6) - Dosimat: 6.2139.000.
Important: A pulse (e.g. a pulse from monitoring or an
activate pulse in a submethod) can set an active output
to inactive! At the end of the TIP method, the outputs
are set to "inactive".
active,
inactive,
pulse, off
info
Message in the display .
The TIP sequence is held and the message displayed.
Continue the sequence with
up to 16
characters
prep
Preparation of titrating burets.
internal D0,
external D1,
external D2
stirrer
Switching stirrer ON/OFF.
In TIP the stirrer is not switched automatically in the
submethods. At the end of TIP the stirrer is switched off
(if stirrer control is ON).
ON, OFF
The parameters of the sequence can be viewed and changed at any time under the
key.