Metrohm 751 GPD Titrino User Manual
Page 126
3.2 Remote control commands
751 GPD Titrino
123
*Parameter Tree part "Parameters for STAT", continuation
Ã
.
Evaluation
Evaluation
³ Ã
.TimeWin
Time windows for rate evaluation
³ ³ Ã
.1
up to 9 windows
³ ³ ³ Ã
.LowLim
Lower limit window 1
0...999 999, OFF
3.2.2.44.
³ ³ ³ Ã
.UpLim
Upper limit window 1
0...999 999, OFF
ditto
³ Ã
.FixVolFix volumes
³ ³ Ã
.1
up to 9 fix volumes
³ ³ ³ Ã
.Value
Value for fix volume 1
0...999 999, OFF
3.2.2.45.
³ Ã
.FixTime
Fix times
³ ³ Ã
.1
up to 9 fix times
³ ³ ³ Ã
.Value
Value for fix time 1
0.01...1, OFF
3.2.2.46.
³ ³
Ã
.
Monitoring
Monitoring
³ Ã
.MeasVal
Monitoring of measured values
³ ³ Ã
.Status
Status
ON, OFF
3.2.2.47.
³ ³ Ã
.LowLim
Lower limit
depends on meas.quant.
ditto
³ ³ Ã
.UpLim
Upper limit
depends on meas.quant.
ditto
³ ³ Ã
.Action
Action if out of limits
end, hold, wait, none
ditto
³ Ã
.Rate
Monitoring of rates
³ ³ Ã
.Status
Status
ON, OFF
3.2.2.48.
³ ³ Ã
.LowLim
Lower limit
0.000...150
ditto
³ ³ Ã
.UpLim
Upper limit
0.000...150
ditto
³ ³ Ã
.Action
Action if out of limits
end, hold, wait, none
ditto
³ Ã
.Temp Monitoring of temperatures
³ ³ Ã
.Status
Status
ON, OFF
3.2.2.49.
³ ³ Ã
.LowLim
Lower limit
-170.0...500.0
ditto
³ ³ Ã
.UpLim
Upper limit
-170.0...500.0
ditto
³ ³ Ã
.Action
Action if out of limits
end, hold, wait, none
ditto
³ Ã
.L10Output
Assignment of output L10
³ ³ Ã
.AssignOutput
To type of limit violation
meas,temp,rate,all,none 3.2.2.50.
³ ³ Ã
.Limit
Assign output to violated limit
upper, lower, any
ditto
³ ³ Ã
.Output
Signal to be set on output L10
active, pulse
ditto
³ Ã
.L11Output
as for output L10
³ Ã
.L12Output
as for output L10
³ Ã
.L13Output
as for output L10
³ ³
Ã
.
Presel
Preselections
³ Ã
.IReq
Request of Id's after start
id1, id1&2, all, OFF
3.2.2.26.
³ Ã
.SReq
Request of sample size after start
value, unit, all, OFF
ditto
³ Ã
.LimSmplSize
Limits for sample size
3.2.2.27
³ ³ Ã
.Status
Status of limit control
ON, OFF
ditto
³ ³ Ã
.LoLim
Lower limit
0.0...999 999
ditto
³ ³ Ã
.UpLim
Upper limit
0.0...999 999
ditto
³ Ã
.RateDisp
Display of drift during cond.
ON, OFF
3.2.2.51.
³ Ã
.ActPulse
Output of a pulse
ON, OFF
3.2.2.28.