Metrohm 751 GPD Titrino User Manual
Page 11

2.4. Configuration, key
751 GPD Titrino
8
meas.input: 1
time interval 7 d
time counter 0 d
service: OFF
next service YYYY-MM-DD
system test report: OFF
If "on" has been set:
Measuring input (1, 2, diff.)
Time interval for validation (1...9999 d)
Time counter (0...9999 d)
Counts the number of days since the last time the
counter was reset.
Monitoring the service interval (ON, OFF)
Monitoring is carried out after the Titrino has been
switched on. If the monitoring responds the message
"Service is due" appears. The message vanishes with
If "on" has been set:
Date of next service (YYYY-MM-DD)
System test report printout (ON, OFF)
With "on" the report of the system test is printed out
after the Titrino has been switched on, see also page
225.
>peripheral units
send to COM1: IBM
send to COM2: IBM
man.reports to COM: 1
balance: Sartorius
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at
the Titrino COM1
"Epson", for Epson
"Seiko", e.g. for DPU-414
"Citizen", e.g. for iDP 562 RS
"HP" e.g. for Desk Jet types. Place curves always at the
beginning of a page as you cannot have them over 2
pages.
"IBM" for all printers with IBM character set Table 437
and IBM graphics, as well as for the data transmission
to a computer or a data system.
COM of Titrino for the output of manually triggered
reports (1, 2, 1&2)
Manually triggered reports e.g. with
Exception
outputted on the COM as defined in the method.
Selection of balance (Sartorius, Mettler, Mettler AT,
AND, Precisa)
Sartorius:
Models MP8, MC1
Mettler:
Models AM, PM and balances with 011,
012, and 016 interfaces
Mettler AT: Model AT
AND:
Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320
Precisa:
Models with RS232C interface