Metrohm 751 GPD Titrino User Manual
Page 39

2.6. Parameters, key
751 GPD Titrino
36
pause 2 0 s
**titr.
extr.time 0 s
**titr.
dos.element: internal D0
I(pol) 1 uA
U(pol) 400 mV
electrode test: OFF
temperature 25.0 °C
cond.
time interval 2 s
cond.
Waiting time after start volume (0...999 999 s)
Second waiting time, e.g. reaction time after dosing of
a start volume. The waiting time can be aborted with
Extraction time (0...999 999 s)
During the extraction time no dosing occurs but the
titration does not stop. The extraction time can be
aborted with
Selection of dosing element (internal D0, external D1/
D2)
internal D0: internal buret of the Titrino
external D1/2: Buret D1, D2, resp.
Polarization current (-127...127 uA)
or the
polarization potential (-1270...1270 mV, in steps of 10
mV) is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby state to a
measurement. "OFF" means that the test is not
performed.
Titration temperature (-170.0...500.0°C).
The temperature is measured at the start of the titration
if a T sensor is connected.
Time interval (1...999 999 s)
Time interval for acquisition of a measured value into
the measuring point list.