4.2. Diagnosis
751 GPD Titrino
202
•
Press
The internal dosing unit expells twice the volume of 3.5 ml.
•
Remove exchange unit.
•
Check spindle zero position, see Fig. 6.
0.2 - 0.6 mm
Fig. 6
Fig. 7
The spindle must be 0.2 - 0.6 mm below the edge of the sliding plate.
The bar of the stopcock coupling must be exactly parallel to the lateral edges of the
GPD Titrino, see Fig.7.
•
Reinsert Exchange Unit.
Titrino fills.
The display of before reappears.
•
Knob 'dV/dt' to right stop.
•
Press the
time measure the time from start to end.
XXX X
********
cylinder empty !
mind selected language!
Spindle remains at maximum position. The transit time of the spindle is
20 s.
•
Measure spindle lifting (can be performed only if the 3.496.0070 Dummy Exchange Unit is
inserted or the locking switch (in right hole) is carefully operated with a screwdriver after
removal of the Exchange Unit).
From the start point, the spindle travels 80 mm. Instead of the spindle height, the ex-
pelled volume can be measured (corresponds to max. vol. of Exchange Unit used).
•
Press
is again in the 'ready' position.
Times for filling:
per stop cock cycle 1 s
for filling
20 s (Tolerance: 10 %)
The following generally holds: Spindle and stopcock must move at a constant speed
(noise!). In the filling setting, the stopcock coupling must position the lever of the
Exchange Unit correctly at the left stop (with virtually no play and without sticking).
•
Set potentiometer 'dV/dt' to left stop.
•
Press
1
/
10 of the cylinder volume to be expelled. The time should be ca. 76 ... 126 s.
•
Set potentiometer 'dV/dt' to right stop.
•
Press