Metrohm 751 GPD Titrino User Manual
Page 156

3.2 Remote control commands
751 GPD Titrino
153
wait:
Interrupt dosing until the limit value is no longer violated, then con-
tinue automatically.
3.2.2.
3.2.2.50.
50. M
Mode.P
.Parameter.M
.Monitoring.L
.L10Output.A
.AssignOutput
meas, temp, rate, all, none
none
M
Mode.P
.Parameter.M
.Monitoring.L
.L10Output.L
.Limit
upper,lower,any
any
M
Mode.P
.Parameter.M
.Monitoring.L
.L10Output.O
.Output
active, pulse
pulse
Parameters for STAT, DOS, DOC: Output of a signal on L10 output (pin 8) of
the remote socket on limit value violation.
The output can be assigned a definite or all monitoring value(s) as well as the
violated limit and the type of signal to be outputted determined.
3.2.2.
3.2.2.51.
51. M
Mode.P
.Parameter.P
.Presel.R
.RateDisp
ON, OFF
OFF
Parameters for STAT, DOC: Display of the current rate during the determination.
3.2.2.
3.2.2.52.
52. M
Mode.P
.Parameter.D
.DosPara.T
.Type
volume
volume, time, rate
M
Mode.P
.Parameter.D
.DosPara.V
.Volume.V
.Volume 0.000...10
10...99999.99
M
Mode.P
.Parameter.D
.DosPara.V
.Volume.D
.DisType
time, rate
rate
M
Mode.P
.Parameter.D
.DosPara.V
.Volume.R
.Rate
0.001...150, max.
max.
M
Mode.P
.Parameter.D
.DosPara.V
.Volume.T
.Time
1...100
100...999 999
M
Mode.P
.Parameter.D
.DosPara.Ti
.Time.T
.Time
1...100
100...999 999
M
Mode.P
.Parameter.D
.DosPara.Ti
.Time.D
.DisType
volume
volume, rate
M
Mode.P
.Parameter.D
.DosPara.Ti
.Time.R
.Rate
0.001...150, max.
max.
M
Mode.P
.Parameter.D
.DosPara.Ti
.Time.V
.Volume
0.000...10
10...99999.99
M
Mode.P
.Parameter.D
.DosPara.R
.Rate.R
.Rate
0.001...150, max.
max.
M
Mode.P
.Parameter.D
.DosPara.R
.Rate.S
.StopType
volume
volume, time
M
Mode.P
.Parameter.D
.DosPara.R
.Rate.T
.Time
1...100
100...999 999
M
Mode.P
.Parameter.D
.DosPara.R
.Rate.V
.Volume
0.000...10
10...99999.99
Parameters for DOS: In the equation Rate = Volume/Time, two parameters are
preset, the third is calculated. Depending on the entry of the type, the following
information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.
3.2.2.
3.2.2.53.
53. M
Mode.P
.Parameter.D
.DosPara.P
.Pause
00...999 999
Parameters for DOS: Pause time in s.
3.2.2.
3.2.2.54.
54. M
Mode.P
.Parameter.D
.DosPara.TD
.TDelta
1...10
10...999 999
Parameters for DOS, DOC: Time interval for the entry of a measured point in the
list of measured points.
3.2.2.
3.2.2.55.
55. M
Mode.P
.Parameter.D
.DosPara.Te
.Temp
-170.0...25.0
25.0...500.0
Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.