beautypg.com

Metrohm 751 GPD Titrino User Manual

Page 155

background image

3.2. Remote control commands

751 GPD Titrino

152

Parameters for STAT: Interpolation of the volume at preset times. Entry of the
times in s. The interpolated volumes are available as variables C5X.

3.2.2.

3.2.2.46.

46. M

Mode.P

.Parameter.E

.Evaluation.FixT

.FixTime.1.V

.1.Value

0.01...1.00, OFF

OFF

Parameters for STAT: Interpolation of the time at preset fraction of the final vol-
ume. Entry as part of V(tot). The interpolated volumes are available as variable
C6X.

3.2.2.

3.2.2.47.

47. M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.S

.Status

ON, OFF

OFF

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.L

.LowLim pH:

0...

±

20.00

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.U

.UpLim U, Ipol: 0...

±

2000

Upol: 0...

±

200.0

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.A

.Action end,hold,wait,none

none

Parameters for STAT, DOC: Monitoring of limit values. Measured values (pH, U
or I) are entered in the list of measured points only if the monitoring is active.
On limit value violation, an action can be selected:
end:

Abort of the determination.

hold:

Interrupt dosing until the sequence is continued by manual switch-
ing (&m $C).

wait:

Interrupt dosing until the limit value is no longer violated, then con-
tinue automatically.

3.2.2.

3.2.2.48.

48. M

Mode.P

.Parameter.M

.Monitoring.R

.Rate.S

.Status

ON, OFF

OFF

M

Mode.P

.Parameter.M

.Monitoring.R

.Rate.L

.LowLim

0.000

0.000...150

M

Mode.P

.Parameter.M

.Monitoring.R

.Rate.U

.UpLim

0.000...150

150

M

Mode.P

.Parameter.M

.Monitoring.R

.Rate.A

.Action

end, hold, wait, none

none

Parameters for STAT: Monitoring of limit values of the rate. Entry in ml/min.
On limit value violation, an action can be selected:
end:

Abort of the determination.

hold:

Interrupt dosing until the sequence is continued by manual switch-
ing (& m $C).

wait:

Interrupt dosing until the limit value is no longer violated, then con-
tinue automatically. On violation of the lower limit, the sequence
can no longer be continued as the rate would become even less
than that when dosing was interrupted.

3.2.2.

3.2.2.49.

49. M

Mode.P

.Parameter.M

.Monitoring.T

.Temp.S

.Status

ON, OFF

OFF

M

Mode.P

.Parameter.M

.Monitoring.T

.Temp.L

.LowLim

-170.0

-170.0...500.0

M

Mode.P

.Parameter.M

.Monitoring.T

.Temp.U

.UpLim

-170.0...500.0

500.0

M

Mode.P

.Parameter.M

.Monitoring.T

.Temp.A

.Action

end, hold, wait, none

none

Parameters for STAT, DOS, DOC: Monitoring of the limit values of the tem-
perature in °C. Temperature values are entered in the list of measured points
only if the monitoring is active.
On limit value violation, an action can be selected:
end:

Abort of the determination.

hold:

Interrupt dosing until the sequence is continued by manual switch-
ing (&m $C).