Metrohm 751 GPD Titrino User Manual
Page 23
2.6. Parameters, key
751 GPD Titrino
20
Titration sequence of DET and MET
After the start, the activate pulse is outputted and the stir-
rer switched on.
The start delay time is waited off.
The sample identifications and the sample size are re-
quested.
The start volume is dispensed (no meas.value acquisition)
and the pause is waited off.
During the titration the volume increments are dispensed
and after each increment a measured value is acquired.
Meas.values are either acquired drift controlled
("equilibrium titration") or a after a fixed waiting time. If you
have not intentionally changed the equilibration time, it is
calculated according to the formula:
150
equilibr.time in s = —————— + 5
ûDrift + 0.01
The criterion (drift or equilibration time) which is first met
applies. This avoids "infinite" titrations. If the parameter
"signal drift" is set "OFF", the measured values are ac-
quired after a fixed equilibration time.
The titration is terminated according to the first criterion
which is met.
The stirrer is switched off.
Evaluations and calculations are carried out.
Data are outputted.
(Activate pulse)
(Stirrer ON)
(Start delay)
(Start conditions)
(Request ident.)
(Request smpl size)
Titration:
Dispense increments
Acquire meas.values
Stop conditions
Calculations
(Stirrer OFF)
Data output