Metrohm 751 GPD Titrino User Manual
Page 147

3.2. Remote control commands
751 GPD Titrino
144
Entry of formulas. Rules for formula entry, see page 74.
Example: "(EP2-EP1)*C01/C00"
In addition to the formula, a text for result output, the number of decimal places
and a unit for the result output can be selected. "No unit" is selected with the
blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is out-
putted in the report full, short, scalc full and scalc srt. It is used for the result
and the corresponding mean value.
The limit control for results can also be activated. If a result is out of limit, a
message appears in the result report, E196 is sent, and output line L13 can be
set.
3.2.2.
3.2.2.6.
6.
M
Mode.Def.S
.Def.SiloCalc.A
.Assign.C
.C24
RSX, EPX, CXX
M
Mode.Def.S
.Def.SiloCalc.A
.Assign.C25
.C25
RSX, EPX, CXX
M
Mode.Def.S
.Def.SiloCalc.M
.MatchId
id1, id1&2, all, OFF
OFF
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
Indication which sample identification(s) have to match so that
the results can be combined.
3.2.2.
3.2.2.7.
7.
M
Mode.Def.C
.Def.ComVar.C
.C30
RSX, MNX, EPX, CXX
M
Mode.Def.C
.Def.ComVar.C31
.C31
etc., up to .C39
.C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.109.
3.2.2.
3.2.2.8.
8.
M
Mode.Def.R
.Def.Report.A
.Assign1
M
Mode.Def.R
.Def.Report.Assign2
.Assign2
DET: full, short, mplist, curve, derive comb, scalc full, scalc srt,
calc, param, calib, ff
MET, SET, MEAS: full, short, mplist, curve, scalc full, scalc srt,
calc, param, calib, ff
KFT: full, short, mplist, curve, scalc full, scalc srt, calc, param, ff
CAL: full, short, scalc full, scalc srt, calc, param, calib, ff
STAT, DOC: full, short, mplist, curve, meas crv, temp crv,
scalc full, scalc srt, calc, param, calib, ff
DOS: full, short,mplist, curve, meas crv, temp crv, scalc full, scalc
srt, calc, param, ff
TIP: full, short, scalc full, scalc srt, calc, param, ff
Definition of the report sequence, which is outputted automatically at the end of
the determination. Entries of more than one block have to be separated with ";".
.Assign1: Output to COM1 of the Titrino. Identical for COM2.
3.2.2.
3.2.2.9.
9.
M
Mode.Def.M
.Def.Mean.1.A
.1.Assign
RS1
RS1, RSX, EPX, CXX
M
Mode.Def.M
.Def.Mean.2.A
.2.Assign
etc., up to .9
.9