4.1 Error and special messages
751 GPD Titrino
191
system error 14
No communication between the Titrino and the connected
Remote Box.
Possible causes:
. The Remote Box was connected when the Titrino was
running
. Titrino has a fault.
. Remote Box has a fault.
Remedy: Set under
Box: OFF", switch off Titrino, take away Remote Box and
switch on Titrino. Contact Metrohm service.
temp out of limit
The determination has been stopped as a limit value of the
temperature monitoring has been violated.
temp
Message in the measuring point list: A limit value of the
temperature monitoring has been violated.
time-out PC keyboard
A connected PC keyboard has been used to call up an ad-
dress (e.g.
rupted.
Possible causes:
. Remote Box has a fault.
. PC keyboard has a fault.
Exit: Correct fault and switch Titrino off/on.
TIP terminated
TIP has been terminated.
transmission error
With a Remote Box connected characters are received which
cannot be interpreted.
Possible causes:
. Wrong key combination has been pressed.
. Wrong PC keyboard has been selected.
. The barcode reader supplies garbled characters.
. The Remote Box has a fault.
Exit: Rectify fault and switch Titrino off/on.
validate instrument
Validation interval has elapsed.
Exit:
wrong card
The card was inserted/removed during an operation with
Exit:
wrong sample
With SET, KFT or DOC with preset titration direction the first
measured value is outside the end point.