3.2. Remote control commands
751 GPD Titrino
176
3.2.2.
3.2.2.164.
164. Se
Setup.M
Mode.S
.StartWait
ON, OFF
OFF
Se
Setup.M
Mode.F
.FinWait
ON, OFF
OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP
(holding point after AutoInfo !".T.GC").
.FinWait:
Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").
3.2.2.
3.2.2.165.
165. Se
Setup.S
.SendMeas.S
.SendStatus
ON, OFF
OFF
Se
Setup.S
.SendMeas.I.Interval
0.08...44...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
3.2.2.167 and 3.2.2.168) in the inputted interval is active.
.Interval:
Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.167 and 3.2.2.168.
The inputted value is rounded off to a multiple of 0.08. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the interface
and on the type of device connection. With "MPList" the measured
values are sent at the time of their entry into the measured point list.
The automatic transmission is switched on/off with 'SendStatus'.
3.2.2.
3.2.2.166.
166. Se
Setup.S
.SendMeas.Sel
.Select
Assembly, Titrator
Titrator
Selection of the unit of which the measured values should be sent (3.2.2.167 or
3.2.2.168).
3.2.2.
3.2.2.167.
167. Se
Setup.S
.SendMeas.A
.Assembly.C
.CyclNo
ON, OFF
OFF
Se
Setup.S
.SendMeas.A
.Assembly.V
.V
ON, OFF
OFF
Se
Setup.S
.SendMeas.A
.Assembly.M
.Meas
ON, OFF
OFF
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.165):
.CyclNo:
Cycle number of the potential measurement. Together with the cy-
cle time (3.2.2.149), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it
is always incremented as long as the instrument remains switched
on.
.V:
Volume
.Meas:
Measured value associated to the cycle number.
The unit "assembly" must be preset (see 3.2.2.166).
3.2.2.
3.2.2.168.
168. Se
Setup.S
.SendMeas.T
.Titrator.C
.CyclNo
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.V
.V
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.M
.Meas
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.d
.dVdt
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.dM
.dMeasdt
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.dMeasdV
.dMeasdV
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.E
.ERC
ON, OFF
OFF
Se
Setup.S
.SendMeas.T
.Titrator.T
.T
ON, OFF
OFF