3.2 Remote control commands
751 GPD Titrino
175
Selects the Titrino COM for the output of automatic info:
&Setup.Keycode
&Setup.Trace
&Setup.SendMeas
&Setup.AutoInfo
3.2.2.
3.2.2.160.
160. Se
Setup.K
.Keycode
ON, OFF
OFF
ON means the key code of a key pressed on the Titrino is outputted. The key
code comprises 2 ASCII characters; table of the keys with their code, see page
195. A keystroke of key 11 is sent as follows:
#11
The beginning of the message is marked by a space (ASCII 32).
3.2.2.
3.2.2.161.
161. Se
Setup.TTree.SShort
ON, OFF
OFF
Se
Setup.TTree.CChangedOnly
ON, OFF
OFF
Definition of the type of answer to $Q.
.Short:
With "ON", each path is sent with only the necessary amount of
characters in order to be unequivocal (printed in bold in this manual). A combi-
nation of .Short and .ChangedOnly is not possible.
.ChangedOnly:
Sends only the changed values, i.e. values which have been
edited. All paths are sent absolute, i.e. from the root.
3.2.2.
3.2.2.162.
162. Se
Setup.Tra
.Trace
ON, OFF
OFF
The Titrino automatically reports when a value has been confirmed with
&SmplData.OFFSilo.Id1"Trace"
The beginning of the message is marked by a space (ASCII 32).
3.2.2.
3.2.2.163.
163. SSetup.L
.Lock.K
.Keyboard
ON, OFF
OFF
Se
Setup.L
.Lock.C
.Config
ON, OFF
OFF
Se
Setup.L
.Lock.P
.Parameter
ON, OFF
OFF
Se
Setup.L
.Lock.S
.SmplData
ON, OFF
OFF
Se
Setup.L
.Lock.U
.UserMeth.R
.Recall
ON, OFF
OFF
Se
Setup.L
.Lock.U
.UserMeth.S
.Store
ON, OFF
OFF
Se
Setup.L
.Lock.U
.UserMeth.D
.Delete
ON, OFF
OFF
Se
Setup.L
.Lock.D
.Display
ON, OFF
OFF
ON means disable the corresponding function:
.Keyboard: Disable all keys of the Titrinos
.Config:
Disable the
.Parameter:Disable the
.SmplData: Disable the
.UserMeth.Recall: Disable "recall" in
.UserMeth.Store: Disable "store" in
.UserMeth.Delete: Disable "delete" in
.Display: Disable the display, i.e. it will not be written to by the device pro-
gram of the Titrino and can be operated from the computer.