7.5 814/815 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
876
■■■■■■■■
viva 1.1
Number of positions
Shows the number of positions on the rack. If no rack is in position, then
the display will be empty.
Shift rate
Sample rack shift rate for manual operation.
Input range
3 - 20 °/s
Default value
20 °/s
[Rack data]
Opens the Rack data dialog window (see Chapter 6.12.3, page 834), in
which the data of the attached rack can be viewed and edited.
[Initialize rack]
Initializes the attached rack. This resets the rack, the lift and the robotic
arm, reads out the rack code and transfers the corresponding rack data to
the Sample Processor.
7.5.5
Properties - Starting position
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Starting position
On the Starting position tab, a position can be defined which is auto-
matically approached when the device is initialized.
Move to starting position
on | off (Default value: off)
If this check box is activated, the position defined here is approached
when the device is initialized.
Tower
Selection of the tower on the sample changer for moving to the required
position.
Only editable for Move to starting position = on and if the corre-
sponding sample changer has two towers, otherwise inactive.
Selection
1 | 2
Default value
1
Move
Selection of the target position which is to be approached:
Only editable for Move to starting position = on, otherwise inactive.
Selection
Rack position | Special beaker | External posi-
tion
Default value
Rack position