6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
762
■■■■■■■■
viva 1.1
Selection
Bit pattern with exactly 14 characters (0, 1, *,
p) | **************
Default value
**************
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
NOTE
We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line statuses.
Pulse length
Duration of the transmitted pulse.
Input range
1 - 999 s
Default value
1 s
Templates for controlling the 843 Pump Station via Remote
Box
The following templates for controlling the 843 Pump Station for draining
and rinsing the measuring vessel are already present and can be neither
edited nor deleted:
Signal
name
Output signal
Pulse
length
Function
Drain
****p*********
25 s
Switching on pump 1 for
25 s (draining)
Drain off
****0*********
1 s
Switching off pump 1
(draining off)
Drain on
****1*********
1 s
Switching on pump 1
(draining on)
Rinse
***p**********
8 s
Switching on pump 2 for
8 s (rinsing)
Rinse off
***0**********
1 s
Switching off pump 2
(rinsing off)
Rinse on
***1**********
1 s
Switching on pump 2
(rinsing on)
These templates for the output lines can be copied and pasted again.
Templates created by the user can be edited and deleted.