2 meas t - general/hardware – Metrohm viva 1.1 User Manual
Page 558
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
546
■■■■■■■■
viva 1.1
Command variables
The following command variables are generated by the command MEAS
T in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identifi-
cation
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has never
been started
DBL
Total duration for the processing of the command in s
ETE
End temperature (temperature after the command has
been processed) in °C
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has never
been started
ITE
Initial temperature (temperature before start conditions are
processed) in °C
NMP
Number of measuring points in measuring point list
STY
Type of stop with which the command was stopped: 1 =
finished normally, 0 = stopped manually with [Quit]
or [Stop]
5.5.2.7.7.2
MEAS T - General/Hardware
Tab: Method
▶ MEAS T ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
1 - 25 characters
The general parameters for the instrument, the sensor and the stirrer are
defined on this tab.