5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
558
■■■■■■■■
viva 1.1
Mode of operation
At the time of the command, the statuses of all of the digital inputs are
entered and saved in command variables (syntax 'Command name.Vari-
able identification')
Parameters
The parameters for the DIGITAL IN command are set on the following
tabs:
■
General
Parameters for instrument.
■
Stop criteria
Parameters for the definition of the stop criteria.
Command variables
The following command variables are generated by the DIGITAL IN com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Port name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
[VAL]
Value of the command variables (optional, i.e. 'Com-
mand name.Portname' = 'Command
name.Portname.VAL'). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited; 1 = active, 0 = inactive
5.5.2.8.3.2
DIGITAL IN - General
Tab: Method
▶ DIGITAL IN ▶ Properties... ▶ DIGITAL IN - 'Command
name'
▶ General
Command name
Name of the command.
Entry
25 characters