2 loop - properties – Metrohm viva 1.1 User Manual
Page 612

5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
600
■■■■■■■■
viva 1.1
Identification
Description
STY
Type of stop with which the LOOP command was
stopped:
0 = manual or after error
1 = maximum number of runs
2 = maximum run time
3 = signal ratio
4 = condition
5.5.2.10.3.2
LOOP - Properties
Dialog window: Method
▶ LOOP ▶ Properties... ▶ LOOP - 'Command name'
Command name
Name of the command.
Entry
25 characters
Stop criteria
Definition of three possible stop criteria. The stop criterion that is reached
first leads to the cancellation of the loop.
Max. run number
on | off (Default value: on)
If this check box is activated, the maximum number of runs is used as a
stop criterion for Repeat and While loops. If the maximum number of runs
has been reached, the loop will be canceled the next time the loop com-
mand is called.
Input range
0 - 999
Default value
2
Maximum run time
on | off (Default value: off)
If this check box is activated, the maximum run time is used as a stop cri-
terion for Repeat and While loops. The starting point for the run time is
the first start of the loop command, i.e., the run time for While loops does
not start until after the first loop run has been started. If the maximum run
time has been reached, the loop will be canceled the next time the loop
command is called.
Input range
0 - 999.9
Default value
0