Metrohm viva 1.1 User Manual
Page 534

5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
522
■■■■■■■■
viva 1.1
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
–1E+12 - 1E+12 (max. 10 digits) mAU
Default value
0 mAU
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are acquired and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Input range
0 - 100 pixels
Default value
0 pixels
Measuring parameters for the 089 Photometer
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
-4,000 - 4,000 mAU
Default value
0 mAU
Measurement
Two types of measurement can be selected, Measurement with drift
control or Measurement without drift control.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control