Measurements

55

Channel Power Measurement

Channel Power measures the Integration Bandwidth, the total power
over a frequency range, concentrated on the center frequency of the
sweep. It is useful for channelized (frequency-division multiplexed)
signals. Results are shown in both total power in the channel (in dBm
or Watts), and spectral density (dBm or Watts per Hz).

Setting Channel Power

Note:

For best accuracy, set the center frequency so the signal

is roughly centered before taking measurements. In addition, the
span should be 1½ – 5 times as large as the desired channel width.

1. Select the Integration Bandwidth (IBW).

Note:

The IBW is automatically set when a channelized

frequency band is selected from the Frequency list.

2. Enter the desired bandwidth.
3. Select the desired unit.

Note:

The default unit is MHz.

Note:

The integration bandwidth, channel power, and

power density will be displayed in the “measurement settings”

area at the bottom of the display. Markers 1 and 2 will be

turned on, and placed at the band edges.

Figure 40 Example, Channel Power

STEP 1

Measurement
Settings