2.3.3.5.5 Rate Ramps Tab
Click the Rate Ramps tab to display the Rate Ramp parameters (see
).
Rate Ramps allow for more than one deposition rate during the deposition phase.
Each Layer of a Process can have an unlimited number of Rate Ramps. Each Rate
Ramp consists of a Start Thickness value that initiates ramping to the Setpoint
value of a new rate, and a Ramp Time value for the time required to ramp to the
new rate Setpoint value.
Figure 2-23 Process window - Rate Ramps tab
Start Thickness (kÅ) . . . . . 0.000 to 999.900 kÅ
The value that, when reached, initiates a timed ramp
to a new rate. The Start Thickness value should be
greater for each subsequent ramp and less than the
Final Thickness of the Layer; otherwise, the Rate
Ramp is ignored.
Ramp Time (sec). . . . . . . . . 0.00 to 30000 s
The time in seconds required to achieve the new rate
Setpoint value.
Setpoint (Å/s) . . . . . . . . . . . -999.90 to 999.90 Å/s
Setpoint (Å/s) is displayed instead of Setpoint % Pwr
when Sensor(s) is selected in the Input box on the
Layer tab. Setpoint (Å/s) is the value of a new rate.