INFICON IQM-233 Thin Film Deposition Controller PCI-Express Card Operating Manual User Manual
Page 79

5 - 5
PN
07
4-
58
4-
P1
A
IQM-233 Operating Manual
Freq (Hz) displays Fail 
during deposition before 
“normal” life of crystal is 
exceeded.
Crystal is being hit by small 
droplets of molten material 
from the evaporation source.
Use a shutter to shield the 
sensor during source 
conditioning.
Move the crystal farther away 
(at least 25.4 cm (10 in.)) 
from the source.
Damaged crystal or 
deposited material is causing 
stress to crystal.
Replace the crystal.
Use an Alloy crystal if 
appropriate for deposited 
material.
Material build-up on crystal 
holder is partially masking 
the crystal surface.
Clean or replace the crystal 
holder.
Freq (Hz) displays Fail 
during deposition before 
“normal” life of crystal is 
exceeded.
Shutter is partially 
obstructing deposition flux or 
sensor is poorly positioned, 
causing uneven distribution 
of material on crystal.
Visually check crystal for an 
uneven coating, and if 
present, correct shutter or 
sensor positioning problem.
Freq (Hz) displays Fail 
when vacuum chamber is 
opened to air.
Crystal was near the end of 
its life; opening to air causes 
film oxidation, which 
increases film stress.
Replace the crystal.
Excessive moisture 
accumulation on the crystal.
Avoid condensation by 
turning off cooling water to 
sensor before opening the 
vacuum chamber to air, and 
then flow heated water above 
the room’s dew point through 
the sensor when the 
chamber is open.
Table 5-1 Symptom/Cause/Remedy Chart (continued)
SYMPTOM
CAUSE
REMEDY
