INFICON IQM-233 Thin Film Deposition Controller PCI-Express Card Operating Manual User Manual
Page 53

3 - 13
PN
07
4-
58
4-
P1
A
IQM-233 Operating Manual
Sensor Fields
The Sensor 1, Sensor 2, and Sensor 3 fields correspond to the sensor(s)
connected to the SMA connectors on the IQM-233 card, with Sensor 1
corresponding to the SMA connector closest to the D-sub connector (refer to
). Each sensor field contains the following parameters:
Density (g/cm
3
) . . . . . . . . . . . . . . . . 0.4 to 99.99 g/cm
3
Enter the density of the material to be deposited using the corresponding
sensor. See
for material density values. The default value is 1.00. If
the density value is not available for a material, see
Tooling (%) . . . . . . . . . . . . . . . . . . . 0.00 to 999.00
Enter a Tooling value to compensate for differences in the measured deposition
rate and actual deposition rate at the substrate due to the sensor or substrate
geometry (see
). The default value is 100.00. To determine the
Tooling value, see
section 6.3, Determining Tooling, on page 6-2
Figure 3-13 Tooling settings
Z-Ratio . . . . . . . . . . . . . . . . . . . . . . . 0.100 to 9.999
Enter a Z-Ratio value specific to the material being deposited to compensate
for the mechanical elasticity of the material to the quartz crystal. See
for material Z-Ratio values.
If the Z-Ratio of a material is not known, enter the default value of 1.000 for
Z-Ratio. Z-Ratio has a smaller effect when the crystal is relatively new. To
determine Z-Ratio experimentally, see
section 6.4, Determining Z-Ratio, on
.
NOTE: Click OK to save the Sensor Setup parameters.
Substrate
Substrate
Tooling
Over 100%
Tooling
Under 100%
Tooling over 100%
Tooling under 100%
Substrate
Substrate