1 carrying out wipe decontamination, Carrying out wipe decontamination – Eppendorf epMotion 5075 User Manual
Page 99

Maintenance
epMotion
®
5075
English (EN)
99
10.1.1
Carrying out wipe decontamination
Prerequisites
• Decontamination agent
• Deionized water
• Lint-free cloths
• Cotton swabs
Cleaning the device
1. Switch off the device and disconnect it from the mains/power line.
2. Moisten the lint-free cloth and cotton swab with decontamination agent.
3. Clean the surface with a lint-free cloth. Clean hard to access surfaces with cotton swabs.
The surfaces are dampened with a decontamination agent.
4. Allow the decontamination agent to take effect.
5. Wipe the decontamination agent with deionized water.
6. Allow the surfaces to dry.
Cleaning the dispensing tools
7. Pull of the ejector from the single-channel dispensing tool.
Abb. 10-1: Removing the single-channel dispensing tool ejector
Fig. 10-1: Removing the single-channel dispensing tool ejector
8. Clean and decontaminate the dispensing tools.
The wipe decontamination of the dispensing tools is identical to the wipe decontamination of the device.