Operation

epMotion

®

5075

English (EN)

93

3. If required, tension the spring clip holder by hand.

Height adapter, Tip Holder und Reservoir Rack

9.3

Executing an application

9.3.1

Performing a test run

If you are executing an application for the first time, perform a test run.

Check the application as follows:

1. Fill the labware with the correct dispensing quantity of demineralized water or a different test liquid.

2. Equip the epMotion worktable with labware and tools.

3. Activate the optical sensor.

4. Fully execute the application.

5. Check to see if the application was executed without errors.

The dispensing is precise and correct.

No liquid sprays out of the reaction vessels. Contamination is unlikely.

Additional information on optimizing the dispensing can be found in the software operating manual.

4. Place the height adapter, Tip Holder and

Reservoir Rack over the positioning pins.

Details on starting and controlling applications can be found in the software operating
manual.

Perform a test run with a liquid which has physical properties that are similar to the liquid for
the application.