5 interfaces, 6 multicon, 7 additional specifications – Eppendorf epMotion 5075 User Manual
Page 112: 1 optical sensor, 2 carrier, 3 gripper, Interfaces, Multicon, Additional specifications 13.7.1, Optical sensor

Technical data
epMotion
®
5075
English (EN)
112
13.5
Interfaces
Only devices which meet the requirements of IEC 950/EN 60950-1 (UL 1950) standards may be connected
to the interfaces.
13.6
MultiCon
You can find information on the technical data in the operating manual of the MultiCon.
13.7
Additional specifications
13.7.1
Optical sensor
13.7.2
Carrier
13.7.3
Gripper
USB
USB 2.0
Ethernet
Ethernet 100 MBit/s
Optical confocal infrared sensor
Non-contact detection of
• Filling levels
• Labware coding and labware height
• Equipping the racks with pipette tips
Detection conditions
90 ± 3° liquid surface to the optical axis
Minimum fill level 3 mm
X-Y-Z axis positioning accuracy
Systematic error
±0.3 mm
Random error
±0.1 mm
Working space
Width: 700 cm (275.6 in)
Depth: 300 cm (118.1 in)
Height: 200 cm (78.7 in)
Form-fit carrying capacity
≤2 000 g
Force-fit carrying capacity
≤1 200 g