3 equipping the thermomixer, Equipping the thermomixer – Eppendorf epMotion 5075 User Manual
Page 40
Hardware
epMotion
®
5075
English (EN)
40
The thermomixer works in a temperature range of 4 °C – 95 °C. The lowest temperature the thermomixer
reaches is 15 °C below ambient temperature. If the ambient temperature is 20 °C, the thermomixer reaches
at least 5 °C.
Tab. 5-2: Maximum speed for labware on the thermomixer
5.4.3
Equipping the thermomixer
When you use a thermoblock or thermoadapter, plates with uneven bases will be optimally tempered, e.g.,
PCR plates.
Equip the thermomixer as follows:
1. Place the plate on the thermoadapter or thermoblock.
2. Place the plate, with thermoadapter or thermoblock, on the thermomixer.
The plate will be fastened to the thermomixer during the application.
Labware
Maximum speed in rpm
Tubes
1000
DWP 96
1200
DWP 384
1200
MTP 6
1000
MTP 96
2000
MTP 384
2000
PCR 96
2000
PCR 384
2000
Reservoir Rack
Cannot be mixed
Thermorack/Thermoblock
1000
Thermoadapter 96
1000
Thermoadapter 384
1000
Thermorack TMX
1300
PrepRack
1300
Information on setting the thermomixer, and a detailed description of all functions, can be
found in the software instructions.