3 equipping the thermal module, Equipping the thermal module – Eppendorf epMotion 5075 User Manual
Page 38
Hardware
epMotion
®
5075
English (EN)
38
Tab. 5-1: Temperature control duration for labware with thermal module
This information applies at an ambient temperature of 25 °C for the border positions in the labware. The
thermal modules were tempered at the target temperature before the labware was positioned.
5.3.3
Equipping the thermal module
When you use a thermoblock or thermoadapter, plates with uneven bases will be optimally tempered, e.g.,
PCR plates.
Equip the thermal module as follows:
1. Place the plate on the thermoadapter or thermoblock.
2. Place the plate, with thermoadapter or thermoblock, on the thermal module.
Labware on the
thermal module
Water filling
volume per well
in μL
Cooling from
25°C to 4°C
Heating from
25°C to 37°C
Heating from
25°C to 95°C
Temperature
setting in °C
min Temperature
setting in °C
min Temperature
setting in °C
min
Thermal module
without labware
5
2
8
Thermoadapter with
96-well twin.tec
PCR plate skirted
50
3
7
38
5
104
11
Thermoblock with 96-well
twin.tec PCR plate skirted
50
3
6
37
4
101
10
Thermoadapter with
384-well twin.tec
PCR plate skirted
10
3
5
37
5
103
8
Thermoblock with
384-well twin.tec
PCR plate skirted
10
3
5
37
5
98
7
24-unit thermorack with
0.5 mL Safe-Lock tubes
350
3
24
37
16
101
35
24-unit thermorack with
1.5 mL Safe-Lock tubes
1200
3
30
38
18
104
44
24-unit thermorack with
2.0 mL Safe-Lock tubes
1700
3
25
37
17
100
40
Information on setting the thermal module and a detailed description of all functions can be