beautypg.com

3 equipping the thermal module, Equipping the thermal module – Eppendorf epMotion 5075 User Manual

Page 38

background image

Hardware
epMotion

®

5075

English (EN)

38

Tab. 5-1: Temperature control duration for labware with thermal module

This information applies at an ambient temperature of 25 °C for the border positions in the labware. The
thermal modules were tempered at the target temperature before the labware was positioned.

5.3.3

Equipping the thermal module

When you use a thermoblock or thermoadapter, plates with uneven bases will be optimally tempered, e.g.,
PCR plates.

Equip the thermal module as follows:

1. Place the plate on the thermoadapter or thermoblock.

2. Place the plate, with thermoadapter or thermoblock, on the thermal module.

Labware on the
thermal module

Water filling
volume per well
in μL

Cooling from
25°C to 4°C

Heating from
25°C to 37°C

Heating from
25°C to 95°C

Temperature
setting in °C

min Temperature

setting in °C

min Temperature

setting in °C

min

Thermal module 
without labware

5

2

8

Thermoadapter with
96-well twin.tec 
PCR plate skirted

50

3

7

38

5

104

11

Thermoblock with 96-well
twin.tec PCR plate skirted

50

3

6

37

4

101

10

Thermoadapter with
384-well twin.tec 
PCR plate skirted

10

3

5

37

5

103

8

Thermoblock with
384-well twin.tec 
PCR plate skirted

10

3

5

37

5

98

7

24-unit thermorack with
0.5 mL Safe-Lock tubes

350

3

24

37

16

101

35

24-unit thermorack with
1.5 mL Safe-Lock tubes

1200

3

30

38

18

104

44

24-unit thermorack with
2.0 mL Safe-Lock tubes

1700

3

25

37

17

100

40

Information on setting the thermal module and a detailed description of all functions can be