beautypg.com

2 decontamination before shipment, 3 servicing the device, 4 servicing the vacuum unit – Eppendorf epMotion 5075 User Manual

Page 101: 1 replacing the seals of the vacuum unit, 2 replacing the silicone mat of the vacuum lid, Decontamination before shipment, Servicing the device, Servicing the vacuum unit 10.4.1, Replacing the seals of the vacuum unit, Replacing the silicone mat of the vacuum lid

background image

Maintenance

epMotion

®

5075

English (EN)

101

10.2

Decontamination before shipment

10.3

Servicing the device

 Have the UV lamp replaced by an Eppendorf service technician once every 2 years.

 Have the HEPA filter replaced by an Eppendorf service technician once a year.

 Have every dispensing tool serviced after 100,000 full strokes or 200,000 strokes.

 Have the seal of the vacuum unit replaced after 300 applications by an Eppendorf AG service

technician.

10.4

Servicing the vacuum unit

10.4.1

Replacing the seals of the vacuum unit

10.4.2

Replacing the silicone mat of the vacuum lid

1. To remove the old silicone mat, undo the screws on the bottom of the VAC Lid.

2. Remove the silicone mat.

3. Place a new silicone mat on the bottom of the VAC Lid.

4. Insert the screws and tighten.

The screw heads must not sit in the silicone mat.

WARNING! Risk to health from contaminated device

1. Follow the instructions in the decontamination certificate. It is available as a PDF file on

our webpage (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.

The Eppendorf AG service team is available to service and certify your device.

The service provisions include:

• Installation qualification (IQ) and operational qualification (OQ)
• Preventive maintenance
• Calibration and validation

Information on the services offered can be found on our webpage: www.eppendorf.com/
epservices.

NOTICE! Leaks will occur in the vacuum chamber if the seal is not serviced
To ensure that the vacuum chamber functions safely, the seal must be replaced after 300
applications.

 After 300 applications the seal must be replaced by an Eppendorf AG service technician.
 Defective seals must be replaced.