Operation

epMotion

®

5075

English (EN)

95

9.3.3.1 Stopping the application immediately

To stop the application immediately, e.g., at risk of collision, proceed as follows:

 Open the front hood.

The application will be stopped immediately.

 You can continue or abort the application.

9.3.3.2 Interrupting the application

 Stop the application using the

Stop application

button (refer to the epMotion software operating

manual).

The current movement will be ended. Then the application stops.

 You can continue or abort the application.

9.3.3.3 Continuing the application

Prerequisites

• The equipping of the worktable has not been changed.
• The levels of the vessels has not been changed.
• The carrier has not been moved by hand.

1. Close the safety screen.

2. Continue the application using the

Continue application

button (refer to the epMotion software operating

manual).

9.3.3.4 Aborting the application

An application cannot be continued after it has been aborted.

Prerequisites

• The application was stopped or interrupted.

 Abort the application using the

Abort application

button (refer to the epMotion software operating

manual).

The application will be aborted.

The tool holder returns the tool to its original location.

The carrier travels to the home position.

The epMotion is ready for operation.

 You can start a new application.

NOTICE! Contamination due to multiple use of pipette tips.
When you abort an application with multiple use of pipette tips, the epT.I.P.S. Motion
SafeRacks contain used pipette tips. The epMotion does not detect which pipette tips have
already been used.

 Dispose of epT.I.P.S. Motion SafeRacks immediately after the application has been

aborted.