Maintenance
epMotion
®
5075
English (EN)
97
10
Maintenance
10.1
Decontamination and cleaning
The device is cleaned as described in the Decontamination section.
WARNING! Danger due to flammable and infectious liquids in the waste container.
Pipette tips in the waste container may contain flammable or infectious liquids.
Wear personal protective equipment.
Handle the pipette tips and sample materials from the waste container in accordance with
the material safety data sheets, safety regulations and laboratory guidelines.
WARNING! Danger due to contamination of the LiquidWasteTub.
The outer surface of the LiquidWasteTub may become contaminated by pipette tips in the
waste container.
Wear personal protective equipment.
Decontaminate the outer surface of the LiquidWasteTub.
CAUTION! Risk of injury from mandrels on the gripper arms.
Each gripper arm has 2 sharp mandrels on the inside.
Handle the gripper from above.
Do not touch the gripper arms on the inside.
NOTICE! Material damage from corroding cleaning agents.
Cleaning and decontamination agents may contain corroding substances. Metal surfaces on
the epMotion become damaged by corrosion.
Use the cleaning agents and disinfectants specified in the operating manual.