beautypg.com

4 vac thermo lid, Vac thermo lid – Eppendorf epMotion 5075 User Manual

Page 44

background image

Hardware
epMotion

®

5075

English (EN)

44

5.6.4

Vac Thermo Lid

Abb. 5-14: VAC Thermo Lid

Fig. 5-14: VAC Thermo Lid

The Vac Thermo Lid has a block with a high heat capacity. The Vac Thermo Lid is heated on a thermal
module. On the vacuum unit, air is sucked in through the slits in the Vac Thermo Lid. The air heats up on
the block and flows through the filter plates.

With the Vac Thermo Lid, you can optimize automated nucleic acid preparation. You can minimize the
alcohol in filter plates, the risk of contamination from alcohol is reduced.

The Vac Thermo Lid can be placed on a thermal module and on the vacuum unit.

The gripper transports the Vac Thermo Lid.

The Vac Thermo Lid cannot be placed on the thermomixer.

5.6.4.1 Using the Vac Thermo Lid

1. Place the Vac Thermo Lid on a thermal module.

2. Heat the Vac Thermo Lid to the required temperature before starting the application.

Heating the Vac Thermo Lid to 110 °C takes approx. 30 min.

CAUTION! Burns from the hot base of the Vac Thermo Lid.
When you temper the Vac Thermo Lid, its base becomes hot.

 Wait until the base of the Vac Thermo Lid has cooled down completely.

Do not place a reservoir rack with reservoirs in the transport route of the Vac Thermo Lid.
Reservoir racks with reservoirs collide with the VAC Thermo Lid on the gripper.