Maintenance
epMotion
®
5075
English (EN)
98
Tab. 10-1: Decontamination methods and decontamination agents
Decontamination method
Decontamination agent
For use on
Wipe decontamination
70% (v/v) ethanol
Worktable
Waste container
LiquidWasteTub
Housing
Work surface adapter
Barcode reader
Vacuum unit
Lid, thermo lid and sealing mat of the
vacuum unit
Labware
Tools
3%–4% sodium hypochlorite
solution
Waste container
LiquidWasteTub
Dispensing Tools
Gripper
Vacuum unit
Lid, thermo lid and sealing mat of the
vacuum unit
Elution plate adapter
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder
Spray decontamination
outside the device
70% (v/v) ethanol
Waste box
LiquidWasteTub
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder
UV irradiation
UV lamp of the CleanCap
1 irradiation cycle lasts 15 min
Worktable
Thermomixer
Thermal module
Autoclaving
121 °C
100 kPa overpressure
20 min
Waste box
LiquidWasteTub
Dispensing Tools
Gripper
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder
Lid, thermo lid and sealing mat of the
vacuum unit
Elution plate adapter