beautypg.com

4 completing your work, 1 cleaning up the worktable, 2 switching off the epmotion – Eppendorf epMotion 5075 User Manual

Page 96: Completing your work 9.4.1, Cleaning up the worktable, Switching off the epmotion

background image

Operation
epMotion

®

5075

English (EN)

96

9.4

Completing your work

9.4.1

Cleaning up the worktable

Complete the following steps after the end of the application.

1. Switch off the HEPA filter, if applicable.

2. Seal the vessels.

3. Remove the labware from the worktable.

4. Empty the waste container.

5. If necessary, decontaminate the device and labware (see Decontamination and cleaning on p. 97).

6. Use a paper towel to soak up the condensation water on the thermal module or thermomixer.

9.4.2

Switching off the epMotion

Proceed as follows.

1. Exit epBlue (refer to the epMotion software operating manual).

2. Switch off the MultiCon.

3. Switch off the epMotion using the mains power switch.

WARNING! Danger due to flammable and infectious liquids in the waste container.
Pipette tips in the waste container may contain flammable or infectious liquids.

 Wear personal protective equipment.
 Handle the pipette tips and sample materials from the waste container in accordance with

the material safety data sheets, safety regulations and laboratory guidelines.

WARNING! Danger due to contamination of the LiquidWasteTub.
The outer surface of the LiquidWasteTub may become contaminated by pipette tips in the
waste container.

 Wear personal protective equipment.
 Decontaminate the outer surface of the LiquidWasteTub.