beautypg.com

2 the device settings tab, The device settings tab, Figure 40: the device settings tab – Kramer Electronics WP-501 User Manual

Page 31: Figure 41: device selection dialog box, Table 6: connect dialog box, On 5.2, Figure 42, A nd, Table 7

background image

KRAMER: SIMPLE CREATIVE TECHNOLOGY

Tab Area Settings

24

5.2 The Device Settings Tab

Device setting readout is possible with standalone room controllers and K-NET Master RC or Aux

K-NET devices connected with a K-NET cable to their Master RC, as defined in the Room Control

tree.

Figure 40

shows the device settings tab:

Figure 40: The Device Settings Tab

1. Click Connect To Device:

The Connect window appears (see

Figure 41

).

2. Select the connection method to the standalone controller or Master RC.

Figure 41: Device Selection Dialog Box

Table 6: Connect Dialog Box

Feature

Function

Connection

Method Area

Check to select the connection to the device via the Ethernet, USB or

Serial port

Ethernet Area IP: Type the IP address of the device to which you want to connect.

Port: shows the port number.

Factory Default Address Button: Click to reset the IP address to its

default value

USB Area

Port: select the communication USB port.

Refresh Ports: click to check if there are ports ready to connect on the

Kramer device

Serial

Select Port: select the communication port

U

Figure 42

U

and

U

Table 7

U

define the device settings after connecting to the device

26F

1

:

Figure 42: The Device Settings Tab (Device with Time Server Options Connected)

1

The actual visible settings change according to the device