Technical data – KROHNE OPTISONIC 7300 EN User Manual

Page 98

background image

8

TECHNICAL DATA

98

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Status output / limit switch

Status output / limit switch

Status output / limit switch

Status output / limit switch

Function and settings

Settable as indicator for direction of flow, overflow, error, operating point.

Status and/or control: ON or OFF

Operating data

Basic I/Os

Modular I/Os

Ex-i

Active

-

U

int

= 24 VDC

I ≤ 20 mA

R

L, max

= 47 kΩ

open:

I ≤ 0.05 mA

closed:

U

0, nom

= 24 V at

I = 20 mA

-

Passive

U

ext

≤ 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

R

L, max

= (U

ext

- U

0

) /

I

max

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

U

ext

= 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

R

L, max

= (U

ext

- U

0

) /

I

max

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

-

NAMUR

-

Passive to

EN 60947-5-6

open:

I

nom

= 0.6 mA

closed:

I

nom

= 3.8 mA

Passive to

EN 60947-5-6

open:

I

nom

= 0.43 mA

closed:

I

nom

= 4.5 mA

U

I

= 30 V

I

I

= 100 mA

P

I

= 1 W

C

I

= 10 nF

L

I

= 0 mH