Operation – KROHNE OPTISONIC 7300 EN User Manual
Page 79
OPERATION
6
79
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
C5.1.4
electronic serial no.
Serial no. of the electronic assembly, can not be changed.
C5.1.5
SW.REV.MS
1
st
line: serial number of the I/O circuit board
2
nd
line: software version number
3
rd
line: production date of the circuit board
C5.1.6
Electronic Revision
ER
1
st
line: serial number of the I/O circuit board
2
nd
line: software version number
3
rd
line: production date of the circuit board
C5.2
display
C5.2.1
language
Select: English / Français / Deutsch
C5.2.2
contrast
Adjust display contrast for extreme temperatures.
Setting: -9…+9
Change is immediate!
C5.2.3
default display
Specification of the default display page that is returned to after a
short delay period.
Select: none (the current page is always active) / 1. meas. page (show
this page) / 2. meas. page (show this page) / status page (show only
status messages) / graphic page (trend display of the
1
st
measurement)
C5.2.5
SW.REV.UIS
Software Revision User Interface Software
1
st
line: serial number of the I/O circuit board
2
nd
line: software version number
3
rd
line: production date of the circuit board
C5.3
1. meas. page
stands for:
3 = meas. page 1
C5.4
2. meas. page
4 = meas. page 2
C5. .1
function
Specify number of measured value lines (font size).
Select: one line / two lines / three lines
C5. .2
measurement 1.line
Specify measurement for 1
st
line.
Select: volume flow / corr. volume flow / mass flow / molar mass / flow
speed / velocity of sound / gain / diagnostics 1, 2, 3
C5. .3
range
0…100% of the measurement set in Fct. C5. .2.
0…xx.xx _ _ _ (format and unit depend on the measurement)
C5. .4
limitation
Set lower and upper limit for frequency output before applying the time
constant.
xxx%
Range: -120…+120%
C5. .5
low flow cutoff
Sets output to "0".
x.xxx ± x.xxx %
Range: 0.0…20%
1
st
value = switching point / 2
nd
value = hysteresis
Condition: 2
nd
value ≤ 1
st
value
C5. .6
time constant
Averages measurement
Increasing value improves stability but delays the reaction.
Range: 0.1…100 s