OPERATION

6

81

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

C5.6.3

load settings

Load saved settings.

Select: cancel (exit function without loading) / factory settings (load in

state as delivered) / backup 1 (load data from storage location 1) /

backup 2 (load data from storage location 2) / load sensor data (factory

settings of calibration data)

Query: go on with copy? (can not be undone)

Select: no (exit the function without saving) / yes (load data from the

selected storage location)

C5.6.4

password quick set

Password required to change data in the quick setup menu.

xxxx (password required)

Range 4 digits: 0001…9999

0000 (= to quick setup menu without password)

C5.6.5

password setup

Password required to change data in the setup menu.

0000 (= to quick setup menu without password)

xxxx (password required); range 4 digits: 0001…9999

C5.6.6

GDC IR interface

This function requires an optical GDC adapter connected to the LCD

display.

Break (exit function without connection)

Activate (Interrupts the optical keys)

After 60 seconds pass without a connection being established or after

the adapter is removed, the function is exited and the optical keys are

active again.

C5.7

units

C5.7.1

size

mm; inch

C5.7.2

volume flow

m

3

/d; m

3

/h; m

3

/min; m

3

/s; l/h; l/min; l/s (l = litres); ext. unit selection

(activates option for more units, sequence see below); cf/d; cf/h; cf/m;

cf/s

C5.7.3

ext. unit selection

Active if “ext. unit selection” is selected in C5.7.2.

MMcf/d; Mcf/d; MMcf/h; Mcf/h; free unit (set factor and text in the next

two functions, sequence see below)

C5.7.4

text free unit

Active if "free unit" is selected in C5.7.3.

For text to be specified refer to

Set free units

on page 83.

C5.7.5

[m

3

/s]*factor

Specification of the conversion factor, based on m

3

/s.

For information refer to

Set free units

on page 83.

C5.7.6

corr. volume flow

MMscf/d; Mscf/d; MMscf/h; Mscf/h; scf/d; scf/h; scf/m; scf/s; Nm

3

/d;

Nm

3

/h; free unit (set factor and text in the next two functions,

sequence see below)

C5.7.7

text free unit

Active if "free unit" is selected in C5.7.6.

For text to be specified refer to

Set free units

on page 83.

C5.7.8

[Normal
m

3

/s]*factor

Specification of the conversion factor, based on normal m

3

/s.

For information refer to

Set free units

on page 83.

C5.7.9

mass flow

lb/h; lb/s; t/h; kg/h; kg/s; free unit (set factor and text in the next two

functions, sequence see below)

C5.7.10

text free unit

Active if "free unit" is selected in C5.7.9.

For text to be specified refer to

Set free units

on page 83.

C5.7.11

[kg/s]*factor

Specification of the conversion factor, based on kg/s.

For information refer to

Set free units

on page 83.

C5.7.15

velocity

m/s; ft/s