Operation – KROHNE OPTISONIC 7300 EN User Manual

Page 78

background image

6

OPERATION

78

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

C3.….7

set counter

Set counter … to the desired value.

Select:

break (exit function) / set value (opens the editor to make the entry)

Query: set counter?

Select: no (exit function without setting the value) / yes (sets the

counter and exits the function)

C3.….8

stop counter

Counter … stops and holds the current value.

Select: no (exits the function without stopping the counter) / yes (stops

the counter and exits the function)

C3.….9

start counter

Start counter

Select: no (exits the function without starting the counter) / yes (starts

the counter and exits the function)

C3.….10 information

1

st

line: serial number of the I/O circuit board

2

nd

line: software version number

3

rd

line: production date of the circuit board

C4 I/O HART

C4

I/O HART

Selection / display of the 4 dynamic variables (DV) for HART

®

.

The HART

®

current output (terminal A basic I/Os or terminal C

modular I/Os) always has a fixed link to the primary variables (PV).

Fixed links of the other dynamic variables (1-3) are only possible if

additional analog outputs (current and frequency) are available; if not,

the measurement can be freely selected from the following list: in

Fct. A4.1 "measurement".

…

stands for 1, 2, 3 or 4

X stands for connection terminals A...D.

C4.1

PV is

Current output (primary variable)

C4.2

SV is

(secondary variable)

C4.3

TV is

(third variable)

C4.4

4V is

(fourth variable)

C4.5

HART units

Unit change of DVs (dynamic variables) in the display; usually varied.

Break: return with 8 key

HART display

®

: copies the settings for the units to the settings for DVs

Standard: factory settings for DVs

C4.….1

current output X

Shows the current analog measured value of the linked current output.

The value can not be changed!

C4.….1

frequency output X

Shows the current analog measured value of the linked frequency

output, if present. The value can not be changed!

C4.….1

HART dynamic var.

Measurements of the dynamic variables for HART

®

.

Linear measurements: volume flow / corr. volume flow / mass flow /

molar mass / flow speed / velocity of sound / gain / diagnostics 1,2,3

Digital measurements: counter 1, 2, 3 / operating hours

C5 device

C5.1

device info

C5.1.1

tag

Settable characters (max. 8 digits): A…Z; a…z; 0…9; / - , .

C5.1.2

C number

CG number, non-alterable (input/output versions)

C5.1.3

device serial no.

Serial no. of the system.