OPERATION

6

77

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

C2.….1

mode

Off (control input switched off)

Hold all outputs (hold current values, not display and counters)

Output Y (hold current values)

All outputs to zero (current values = 0%, not display and counters)

Output Y to zero (current value = 0%)

All counters (reset all counters to "0")

Counter "Z" reset (set counter 1, 2 or 3 to "0")

Stop all counters

Stop counter "Z" (stops counter 1, 2 or 3)

Zero outp.+stop Cnt. (all outputs 0%, stop all counters, not the display)

External range Y (control input for external range of current output Y) -

also make this setting on current output Y (no check if current output Y

is available)

Error reset (all resettable errors are deleted)

C2.….2

invert signal

Select: off / on

C2.….3

information

1

st

line: serial number of the I/O circuit board

2

nd

line: software version number

3

rd

line: production date of the circuit board

C2.….4

simulation

Sequence see B 1.… control input X

C3 I/O counter

C3.1

counter 1

Set function of counter …

…

stands for 1, 2, 3

C3.2

counter 2

C3.3

counter 3

C3.….1

funct. of counter

Define counter

Select:

Sum counter (counts positive and negative values)

+counter (counts only the positive values)

-counter (counts only the negative values)

Off (counter is switched off)

C3.….2

measurement

Selection of the measurement for counter …

Select: volume flow / mass flow / corr. volume flow

C3.….3

low flow cutoff

Sets output value to "0".

Range: 0.0…20%

1

st

value = switching point / 2

nd

value = hysteresis

Condition: 2

nd

value ≤ 1

st

value

C3.….4

time constant

Averages measurement

Increasing value improves stability but delays the reaction.

Range: 000.0…100.0 s

C3.….5

preset value

If this value is reached, positive or negative, a signal is generated that

can be used for a status output at which "preset counter X" has to be

set.

Preset value (max. 8 digits) x.xxxxx in selected unit, see C5.7.9 + 12

C3.….6

reset counter

Sequence see Fct. A3.2, A3.3 and A3.4