Electrical connections – KROHNE OPTISONIC 7300 EN User Manual

Page 44

background image

4

ELECTRICAL CONNECTIONS

44

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Pulse and frequency output passive P

N

NAMUR, modular I/O

• Connection in conformity with EN 60947-5-6
• open:

I

nom

= 0.6 mA

closed:
I

nom

= 3.8 mA

• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

INFORMATION!

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.
Wall-mounted version:

Wall-mounted version:

Wall-mounted version:

Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
(insulation to DIN 46245) in the terminal compartment.

Any connection polarity.

Figure 4-15: Pulse and frequency output passive P

N

to NAMUR EN 60947-5-6