8

TECHNICAL DATA

92

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Display and user interface

Display and user interface

Display and user interface

Display and user interface

Graphic display

LC display, backlit white

Size: 128x64 pixels, corresponds to 59x31 mm = 2.32"x1.22"

Display turnable in 90° steps.

The readability of the display could be reduced at ambient temperatures

below -25°C / -13°F.

Operator input elements 4 optical keys for operator control of the signal converter without opening the

housing.

Option: Infrared interface (GDC)

Remote control

PACTware

®

including Device Type Manager (DTM)

All DTM's and drivers will be available at the internet homepage of the

manufacturer.

Display functions

Display functions

Display functions

Display functions

Menu

Programming of parameters at 2 measured value pages, 1 status page,

1 graphic page (measured values and descriptions adjustable as required)

Language of display

texts

English, French, German

Units

Metric, British and US units selectable from list / free unit.

Measuring accuracy

Gas flow (uncorrected)

Gas flow (uncorrected)

Gas flow (uncorrected)

Gas flow (uncorrected)

Reference conditions

(for gas calibration)

Medium: Air

Temperature: 20°C / 68°F

Pressure: 1 bar / 14.5 psi

Theoretical calibration

(standard)

DN100...600 / 4…24": < ± 1.5% of actual measured flow rate, for 1...30 m/s

DN50...80 / 2…3": < ± 3% of actual measured flow rate, for 1...30 m/s

Gas calibration

DN100...600 / 4…24": < ± 1% of actual measured flow rate, for 1...30 m/s

DN50...80 / 2…3": < ± 2% of actual measured flow rate, for 1...30 m/s

Repeatability

< ± 0.2%

Operating conditions

Temperature

Temperature

Temperature

Temperature

Process temperature

Compact version

Compact version

Compact version

Compact version

-40...+125°C / -40...+257°F

-40...+180°C / -40...+356°F, max. ambient temperature of 40°C / 104°F

Remote version

Remote version

Remote version

Remote version

-40...+180°C / -40...+356°F

Compact & remote version

Compact & remote version

Compact & remote version

Compact & remote version

Carbon steel flanges acc. to EN 1092-1,

min. process temperature: -10°C / +14°F

Carbon steel flanges acc. to ASME, min. process temperature: -29°C / -20°F

FFKM transducer o-rings, min. process temperature: -20°C / -4°F

Ambient temperature

Standard (die-cast aluminum converter housing): -40…+65°C / -40…+149°F

Option (die-cast stainless steel converter housing): -40...+55°C / -40...+131°F

Storage temperature

-50…+70°C / -58…+158°F