ELECTRICAL CONNECTIONS

4

51

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Status output/limit switch S

N

NAMUR, Ex i I/Os

• Connection acc. to EN 60947-5-6
• open:

I

nom

= 0.43 mA

closed:
I

nom

= 4.5 mA

• The output is closed when the device is de-energized.
• X designates the connection terminals B or D, depending on the version of the signal

converter.

INFORMATION!

Any connection polarity.

Figure 4-27: Status output/limit switch S

N

to NAMUR EN 60947-5-6 Exi