ELECTRICAL CONNECTIONS

42

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Pulse/frequency output active, modular I/Os

• U

nom

= 24 VDC

• f

max

in the operating menu set to f

max

≤ 100 Hz:

I ≤ 20 mA
open:
I ≤ 0.05 mA
closed:
U

0, nom

= 24 V at I = 20 mA

• f

max

in operating menu set to 100 Hz < f

max

≤ 10 kHz:

I ≤ 20 mA
open:
I ≤ 0.05 mA
closed:
U

0, nom

= 22.5 V at I = 1 mA

U

0, nom

= 21.5 V at I = 10 mA

U

0, nom

= 19 V at I = 20 mA

• If the following maximum load impedance R

L, max

is exceeded, the load impedance R

L

must

be reduced accordingly by parallel connection of R:
f ≤ 100 Hz: R

L, max

= 47 kΩ

f ≤ 1 kHz: R

L, max

= 10 kΩ

f ≤ 10 kHz: R

L, max

= 1 kΩ

• The minimum load impedance R

L, min

is calculated as follows:

R

L, min

= (U

ext

- U

0

) / I

max

• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

INFORMATION!

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.
Wall-mounted version:

Wall-mounted version:

Wall-mounted version:

Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
(insulation to DIN 46245) in the terminal compartment.

Any connection polarity.

Figure 4-13: Pulse / frequency output active P

a