beautypg.com

Electrical connections – KROHNE OPTISONIC 7300 EN User Manual

Page 47

background image

ELECTRICAL CONNECTIONS

4

47

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Control input passive, modular I/Os

• 3 V ≤ U

ext

≤ 32 VDC

• I

max

= 9.5 mA at U

ext

≤ 24 V

I

max

= 9.5 mA at U

ext

≤ 32 V

• Switching point for identifying "contact open or closed":

Contact open (off): U

0

≤ 2.5 V with I

nom

= 1.9 mA

Contact closed (on): U

0

≥ 3 V with I

nom

= 1.9 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

Control input active C

N

NAMUR, modular I/Os

• Connection acc. to EN 60947-5-6
• Switching point for identifying "contact open or closed":

Contact open (off): U

0, nom

= 6.3 V with I

nom

< 1.9 mA

Contact closed (on): U

0, nom

= 6.3 V with I

nom

> 1.9 mA

• Detection of cable break:

U

0

≥ 8.1 V with I ≤ 0.1 mA

• Detection of cable short circuit:

U

0

≤ 1.2 V with I ≥ 6.7 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

Figure 4-20: Control input passive C

p

1 Signal

CAUTION!
Observe connection polarity.

Figure 4-21: Control input active C

N

to NAMUR EN 60947-5-6