OPERATION

68

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

6.3.3 Menu C, setup

No.

No.

No.

No.

Function

Function

Function

Function

Settings / descriptions

Settings / descriptions

Settings / descriptions

Settings / descriptions

C1 process input

C1.1

meter size

Select from size table.

Range: DN25...1000 mm / 1...40"

C1.2

calibration

C1.2.1

zero calibration

Display of current zero calibration value.

Query: calibrate zero?

Setting: cancel / automatic / default

Automatic (shows the current value as the new zero calibration value).

C1.2.2

GK

Select GK value (see nameplate of the measuring sensor).

Range: 0.5000...10.000

C1.3

filter

C1.3.1

limitation

Limitation of all flow values, before smoothing by time constant,

affects all outputs.

Range: -100.0….+100.0 m/s

C1.3.2

flow direction

Define polarity of flow direction.

Select: normal (according to the arrow on the measuring sensor) /

backwards (in the opposite direction to the arrow)

C1.3.3

time constant

Averages measurement

Increasing value improves stability but delays the reaction.

Master setting, overrules time constent settings of all outputs.

xxx.x s

Range: 0.0…100 s

C1.3.4

low flow cutoff

Sets output value of all outputs to zero. "0" appears in display.

x.xxx ± x.xxx m/s

Range: 0.0…10 m/s

1

st

value = switching point / 2

nd

value = hysteresis

Condition: 2

nd

value ≤ 1

st

value

C1.4

plausibility

Changes in values outside error limit are only accepted when number

of measurements exceeds counter limit (C1.4.3).

C1.4.1

error limit

Measurement is ignored if value exceeds the error limit.

One ignored value = counter +1.

Measurements within the error limit decrease the counter

(see C1.4.2).

Range: 0…100%

C1.4.2

counter decrease

Multiplier for every measurement within the error limit to decrease

the counter.

Range: 01…99

C1.4.3

counter limit

Above this value measurements will not be ignored.

Range: 000…999

C1.5

simulation

C1.5.1

volume flow

Query: sim. volume flow.

Select: cancel / set value (select value, start simulation? yes / no).