OPERATION

72

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

C2.….4

error condition

The following error conditions can be selected.

Select: error in device (error category [F]) / application error

(error category [F]) / out of specification (error category [S])

C2.….5

measurement

Measurements for activating the output.

Select: volume flow / corr. volume flow / mass flow / molar mass /

flow speed / velocity of sound / gain / diagnostics 1, 2, 3.

C2.….6

range

0…100% of the measurement set in Fct. C2.….5

0…xx.xx _ _ _ (format and unit depend on the measured variable, see

above)

C2.….7

polarity

Set polarity of current output, please note flow direction in C1.3.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C2.….8

limitation

Set lower and upper limit for current output before applying the time

constant (see Fct. C2.….10).

±

xxx … ±xxx%

Range: -150…+150%

C2.….9

low flow cutoff

Below set value, current output is set to zero.

x.xxx ± x.xxx%

Range: 0.0…20%

1

st

value = switching point / 2

nd

value = hysteresis

Condition: 2

nd

value ≤ 1

st

value

C2.….10 time constant

Averages measurement

Increasing value improves stability but delays the reaction.

Range: 000.1…100.0 s

C2.….11 special function

When activated, changing the scale improves the resolution.

Select:

Off (switched off)

Automatic range (scale changes to extended automatically for the

threshold with hysteresis. Switching from one to another scale

requires activation by a status output)

External range (scale changes to extended range by control input)

C2.….12 threshold

Appears only when Fct. C2.….11 is activated.

Set lag value between normal and extended range. The automatic

range function always changes from the extended to the normal range

when the 100% current is reached.

Range: 05.0…80%

C2.….13 information

1

st

line: serial number of the I/O circuit board

2

nd

line: software version number

3

rd

line: production date of the circuit board

C2.….14 simulation

Sequence see B1.… current output X

C2.….15 4mA trimming

Trimming of the current at 4 mA

Reset to 4 mA restores the factory calibration.

Used for HART

®

setting.

C2.….16 20mA trimming

Trimming of the current at 20 mA

Reset to 20 mA restores the factory calibration.

Used for HART

®

setting.